3. Operating instructions
page 16
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
F
IGURE
:
M
AIN MENU
Following functions can be selected from the main menu:
F1: Start Measuring = measurement cycle starts with automatic sampling, followed by gold
trap heating and atomic absorption measurement.
F2: Set Sample Parameters = set sample volume or measurement cycle duration.
F3: Data Logger = data logger management: switch on/off data logger, display logged data,
transfer logged data to PC, clear logged data.
F4: Calibration = show actual or all calibrations or perform new calibration or span check.
Page2= show next page
F1: Print last data: all data (up to 200 measurements) since the last sample volume setting are
printed. For this function a printer must be connected to the parallel printer output.
F2: Continuous print on/off: results are printed automatically if a measurement is completed.
For this function a printer must be connected to the parallel printer output.
F3: Set 4-20 mA: F1- allows setting of measuring unit µg/m³ - ng/m³; F2- setting of the
measuring range relevant for the 4-20 mA signal. F3- the 4-20 mA output can be tested, the
output signal will be according to the set mA value..