International Electrical Symbols Double Insulated Grounding Warning. Refer to the Operating Manual AC (

Strany 3

10UT207A/208A/209A Operating ManualThe Meter Structure (see gure 1) ① Hand Guards: to protect user’s hand from touching the dangerous area.

Strany 4 - Overview

11UT207A/208A/209A Operating ManualDisplay Symbols (see gure 2)Figure 21. Test of diode2. The continuity buzzer is on3. Indicator for zeroing4. D

Strany 5 - Unpacking Inspection

12UT207A/208A/209A Operating Manual6. True RMS indicator7. Indicator for AC voltage or current8. Indicates negative reading9. Indicator for DC vol

Strany 6 - Safety Information

13UT207A/208A/209A Operating ManualFunctional Buttons Below table indicated for information about the functional button operations.SELECTRANGEHOLDButt

Strany 7 - Rules for Safe Operation

14UT207A/208A/209A Operating ManualZEROWhen the Meter is at %Hz, V and A , press %Hz to measure frequency and duty cycle.Press ZERO to zero

Strany 8

15UT207A/208A/209A Operating ManualThe Effectiveness of Functional ButtonsNot every functional buttons can be used on every rotary switch positions.

Strany 9

16UT207A/208A/209A Operating ManualMeasurement Operation A. DC/AC Voltage Measurement (see gure 3)Figure 3

Strany 10

17UT207A/208A/209A Operating Manual WarningTo avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages h

Strany 11

18UT207A/208A/209A Operating Manual circuit impedance is less than or equal to 10kΩ, the error is negligible (0.1% or less). ● When DC

Strany 12 - Display Symbols (see gure 2)

UT207A/208A/209A Operating Manual13456910111315161618202225273133OverviewUnpacking InspectionSafety Information Rules for Safe OperationInternational

Strany 13

19UT207A/208A/209A Operating ManualFigure 4Note: ● To obtain a more precise reading, you could remove the objects being tested from the cir

Strany 14 - Functional Buttons

20UT207A/208A/209A Operating Manual input terminals beforehand, press ZERO to automatically subtract the value measured when the testing

Strany 15

21UT207A/208A/209A Operating Manual measurement mode.3. The buzzer sounds if the resistance of a circuit under test is less than 10Ω. 4. The buzze

Strany 16

22UT207A/208A/209A Operating ManualNote ● To avoid harms to you or damages to the Meter from eletric shock, do not attempt to input voltages

Strany 17 - Measurement Operation

23UT207A/208A/209A Operating Manual1. Insert the red test lead into the VΩHz terminal and black test lead into the COM terminal.2. Set the rotary

Strany 18

24UT207A/208A/209A Operating ManualFigure 6

Strany 19

25UT207A/208A/209A Operating ManualE. Frequency and Duty Cycle Measurement (see gure 7) WarningTo avoid harms to you or damages to the Meter fr

Strany 20

26UT207A/208A/209A Operating ManualFigure 7

Strany 21

27UT207A/208A/209A Operating ManualF. DC/AC Current Measurement (see gure 8) WarningThe operating temperature must be 0℃ ~40℃ when measuring cu

Strany 22

28UT207A/208A/209A Operating ManualFigure 8

Strany 23

2UT207A/208A/209A Operating Manual35353637434344Specications A. General Specications B. Environmental RequirementsAccuracy Specications

Strany 24

29UT207A/208A/209A Operating Manual zeroing. ● Center the conductor within the transformer jaw, then release the Meter slowly until t

Strany 25 - Figure 6

30UT207A/208A/209A Operating Manual ● The meter will zero automatically. ● When the measuring current >1A, press Hz button to toggle betwe

Strany 26

31UT207A/208A/209A Operating ManualG. Temperature Measurement (UT208A only, see gure 9)To measure temperature measurement, connect the Meter as follo

Strany 27 - Figure 7

32UT207A/208A/209A Operating ManualFigure 9

Strany 28

H. Capacitance measurement (UT208A only, see gure 10) To avoid harms to you or damages to the Meter from eletric shock, do not attempt to inpu

Strany 29 - Figure 8

33 34UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualFigure 10

Strany 30

35 36UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualSpecicationsA.General Specications: ● Display: 3 3/4 digits LCD display, M

Strany 31

35 36UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualB. Environmental Requirements ● The Meter is suitable for indoor use. ● A

Strany 32

37 38UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualAccurate Specications Accuracy: ±(a% reading + b digits), guarantee for 1 year

Strany 33 - Figure 9

37 38UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualB. AC Voltage Range Resolution Accuracy Overload protection

Strany 34

OverviewThis Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings a

Strany 35 - Figure 10

39 40UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualC. Resistance Range Resolution Accuracy Overload prote

Strany 36 - Specications

39 40UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualE. Diode Test Range Resolution Accuracy

Strany 37

41UT207A/208A/209A Operating ManualG. Duty Cycle Range Resolution Accuracy Overload Protection 0.1%~99.9% 0.1

Strany 38

42UT207A/208A/209A Operating ManualJ. Temperature (UT208A only) Range Resolution Accuracy Overload Protection

Strany 39

43 44UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualMaintenance This section provides basic maintenance information including batt

Strany 40

43 44UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualB. Replacing the Battery (see gure 11) WarningTo avoid false readings

Strany 41

45UT207A/208A/209A Operating ManualUT207A/208A/209A Operating ManualMake sure the transformer jaw and the tets leads are disconected from the circuit

Strany 42

45UT207A/208A/209A Operating ManualUT207A/208A/209A Operating Manual

Strany 43

4UT207A/208A/209A Operating ManualUnpacking InspectionOpen the package case and take out the Meter. Check the following items carefully to see any mis

Strany 44 - Maintenance

5Safety InformationThis Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT II 600V, CAT III 300V) and double

Strany 45

6UT207A/208A/209A Operating ManualRules for Safe Operation WarningTo avoid possible electric shock or personal injury, and to avoid possible dam

Strany 46

7UT207A/208A/209A Operating Manual changeover of range shall be made during measurement is conducted to prevent damage of the Meter. ●

Strany 47

8UT207A/208A/209A Operating Manual electrical specications replacement parts. ● The internal circuit of the Meter shall not be altered at wi