UNI-T UT33A Specifikace Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 42
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
Measurement Operation (9)
22
Model UT33A: OPERATING MANUAL
Note
l The test leads can add 0.1to 0.3of error to resistance measurement. To
obtain precision readings in low-resistance measurement, that is the range of
200, short-circuit the input terminals beforehand and record the reading
obtained (called this reading as X). (X) Is the additional resistance from the test
lead. Then use the equation: measured resistance value (Y) - (X) = precision
readings of resistance.
l For high-resistance measurement (>1M), it is normal taking several seconds
to obtain a stable reading.
l When resistance measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
Zobrazit stránku 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 41 42

Komentáře k této Příručce